Trace Element Tubes

0 items in your bag
Quotation Request

contact us

We are readily availbale to get your inquiry:

email: degroot@degroot.co.il phone: +972-3-9039999 fax: +972-3-9039090
X

Vacuette

Catalogue number *
Description
Specification
 
Application
VACUETTE® Trace Element Tubes are used for determinations in serum / heparinised plasma for the determination of trace elements.
NH tubes with royal blue cap are recommended.
 
Materials
VACUETTE® Trace Element Tubes are made out of PET (polyethylenetherephtalate). The caps are produced
from PE (polyethylene) and are coloured. The rubber component of the cap is composed of Brom Butyl Caoutchouc. 
The stabilisation ring is composed of PP 
(polypropylene).
 
Packaging
Tubes are packed in 50 pieces per rack wrapped with a polyethylene foil. Tubes are packed in 24 racks of 50 tubes. Each box is filled with 1200 tubes.
 
Shelf life
depends on fill volume - information upon request
 
General Information / Special aspects
 
The heparin concentration of VACUETTE® Plasma Trace Element Tubes is standardised: 12-30 I.U. (International Units) of sodium heparin per 1ml blood (ISO 6710).
The molecular mass is between 3.000 and 30.000 Daltons. VACUETTE® Plasma Trace Element Tubes are clearly identifiable by their dark blue safety cap.
 
VACUETTE® Trace Element Sodium Heparin Tubes Contamination Upper Limits (ppb)
Ag  Silver 
< 1.0*1
Mn  Manganese 
< 1.0 *1
Al  Aluminium
< 2.6
Mo  Molybdenum
< 1.0*1
As  Arsenic 
< 1.0*1
Ni  Nickel
< 1.0*1
Ba  Barium
< 1.0*1
Pb  Lead 
< 0.2*1
Be  Beryllium 
< 1.0*1
Se  Selenium
< 1.0*1
Bi  Bismuth
< 1.3
Sb  Antimony
< 1.0*1
Cd  Cadmium  
< 0.2*1
Sn  Tin
< 1.0*1
Cr  Chromium 
< 1.0*1
Te  Tellurium
< 1.0*1
Co  Cobalt 
< 1.0*1
Th  Thorium
< 1,0*1
Cu  Copper
< 1.0*1
Tl  Thallium
< 0,2*1
Hg  Mercury 
< 0.2*1
U  Uranium
< 1.0*1
I  Iodine
< 1.0*1
Zn  Zinc
< 2.9
Li  Lithium
< 1.0*1
 
1detection limit of ICP-MS


 
From these results it can be concluded that the VACUETTE®  NH Trace Element Sodium
Heparin tube is suitable for trace element analysis in  blood for the tested trace elements except Aluminium
(Al) and Bismuth (Bi). The contamination level found for Zinc (Zn) is far below the normal range of Zn in whole blood.
1
X

הזמנת לקוחות רשומים

לקבלת טופס על המסך יש להקיש את מספר הלקוח שלך בחברת דה גרוט בע"מ

X

הזמנות ישירות - כללית שרותי בריאות

:לקבלת טופס הזמנה על המסך יש להקיש את הנתונים הבאים

Read more about us

X

contact us

We are readily availbale to get your inquiry:

phone:   +972-3-9039999 fax:   +972-3-9039090